Graphical User Interface demo
is on-line!
|
|
|
|
|
Welcome to Cross Flow Filtration – TFF System page. |
|
|
 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
PendoTECH TFF Process Control System is an optimal solution that adds automation and data collection to a Cross Flow / Tangential Flow Filtration - TFF) Process.
System includes:
- A powerful yet easy to use Graphical User Interface
- Recipe based control of entire concentration and diafiltration process with multiple options for automated diafiltration end point (filtrate weight, conductivity, pH,
air in feed line, total flow)
- Pressure measurement and calculation of trans-membrane pressure (TMP) and delta-P with PendoTECH Pressure Sensors
- Automated recovery recipe with customizable options
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
| |
|
 |
|
|
|
|
- Flow measurement with different flow meter options including flow totalizing (rotary, mag, ultrasonic, Coriolis)
- Process control via integration of industry standard scales & pumps (peristaltic, rotary lobe, diaphragm, etc)
- No process scale limitations
- Conductivity and pH probe inputs with software calibration routines (conductivity transmitter with temperature display)
- Inputs available for other probes/sensors with 4 - 20 milliamp outputs such as UV and temperature
- Accessory Valves for Diafiltration 1 / 2 control
- Pressure alarm set-points that will shut down the system; notification set-point warnings that can generate email alerts
- Optional 2L vessel with low-point drain and stirrer
- Options for TMP control with non-product contact PendoTECH Throttling Pinch Valve
- Ability to control Retentate Flow
- Available with IQ/OQ templates
- PendoTECH can deliver the system with desired components including a turnkey system(CLICK FOR PIC)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
|
|
|
SEE LINKS ABOVE RIGHT FOR PRODUCT PRESENTATION AND SOFTWARE DEMO |
|
|
|
|
|
| |
|
|
|
|
| |
Graphical User Interface |
|
|
|
|
|
Overview of Graphical User Interface (please contact us for an immediate web demo!)
Four tabs are used to navigate the software:
1. Setup View- used to enter experiment information, program recipe, set alarms and to create a data file where all process data will be stored
2. System View- used to view current process values and change pump flow rates
3. Trends View- utility to trend process variables that is loaded with features to dynamically view the data of interest, ability to quickly auto-scale or manually scale the axes, and ability to export a trend of interest either as data or a graphics file
4. Maintenance View- used to set up pumps based on tubing size, calibrate pH and conductivity via easy to use wizards, select units of measure, configure flow meter and more.
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
System View
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
|
 |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |
|
Setup View
Trends View
 Maintenance View
|
|
|
|
|
|
|
Process Schematic with All Inputs and Outputs Shown
|
|
|
|
|
|
 |
|
| |
Back Panel of TFF System with inputs for pressures (3), filtrate flow, 2 scales, air detector for buffer pump, pH, conductivity & 2 spare analog inputs; outputs for 3 pumps; spare holes for additional options that are enabled in software such as recovery valves and diafiltration selection valves:

|
|
|
|
|
|
|
|
|
| |
 |
|
|
|
|
|
|
|
|
|
|
|
Please contact us at request@pendotech.com for more information. Click here to fill out our online form to join our mail list for new product and service updates.
Copyright © 2011, PendoTECH |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|